13
5. Operation
5.1 Operation of the LJ16
Controls
(1) Control bar of the balance (tare bar)
(2) Keypad of the dryer
Switching display on/off
• Press control bar (1) briefly ; all display segments light up for a few seconds.
• The display then shows 0.000 g.
• Raise control bar briefly; the display blanks out.
Taring
• Place aluminum sample pan on pan support.
• Press control bar (1) briefly; the display changes to 0.000 g.
Calibrating
Before work is performed for the first time with the LJ16 dryer, the balance should be calibrated.
To achieve a correct calibration, it is advisable to attach the instrument to the power supply 30
minutes before calibrating. Calibration is necessary only if the absolute sample weight is of
importance.
• Press control bar (1) until -CAL- appears, then release; display changes to - - - - - - and then to
calibration weight (flashes).
• Load requested calibration weight (e.g. 100 g); display changes to - - - - - -, then 0.000 g
flashes.
• Remove calibration weight; display shows - - - - - - and then changes to 0.000 g.
The instrument is now calibrated.
1
2
0.000
9