After the calibration the Multiplier or slope calibration factor ”M” and the Adder or offset calibra-
tion factor ”A” are displayed.
Select Yes to save the calibration values and the Successful Calibration is confirmed on the dis-
play. If an ISM sensor is connected, the calibration data will be stored in the sensor.
You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing
”Enter” the M300 returns to the measuring mode.
7.2.3 ProcessCalibration
Enter Conductivity Sensor Calibration mode as described in section 7.1 “Enter Calibration Mode”
and choose one of the compensation modes (see section 7.2 “Conductivity/Resistivity Calibra-
tion”).
Select Process Calibration and press [ENTER]. With conductitivty sensors a process calibration
is always performed as a slope calibraton.
Take a sample and press the [ENTER] key again to store the current measuring value.
During the ongoing calibration process, the letter of the channel, which is concerned by the cali-
bration, ”A” or “B” is blinking in the display.
After determining the conductivity value of the sample, press the [CAL] key again to proceed
with the calibration.
Enter the conductivity value of the sample, then press the [ENTER] key to start the calculation of
calibration results.
After the calibration the Multiplier or slope calibration factor ”M” and the Adder or offset calibra-
tion factor ”A” are displayed.
Select Yes to save the calibration values and the Successful Calibration is confirmed on the dis-
play.