Continuous conductivity monitoring
according to USP <645>
USP guideline <645> sets a standard
for the quality assessment of USP
waters based on measurement of the
electrolytic conductivity. There is a
3-stage test in which stage 1 allows
on-line, non-temperature compensated
conductivity measurement. There are
specific requirements for the sensors
and transmitters (see table below).
Thornton instruments fulfill all these
requirements. In addition, Thornton
instruments provide USP and EP set-
points for convenience.
METTLER TOLEDO instruments
meet USP <645> water
conductivity requirements
Application guide for conductivity sensors
4-electrode sanitary
Specification USP <645>
Conductivity sensor Verify cell constant within
and cell constant accuracy ± 2 % using a reference solution
Conductivity meter NIST traceable 0.1 % precision
calibration resistors in place of sensor
Instrument resolution 0.1 µS / cm
Instrument accuracy 0.1 µS / cm
at 1.3 µS / cm
Temperature compensation Must be read uncompensated
Instrument dynamic range 10