7 Calibration
(PATH: Cal)
7.1 Exit calibration mode
C
NOTE: to exit Calibration mode at any time press the and key simultaneously (escape).
The transmitter returns to the Measurement mode and the old calibration values remain active.
7.2 Enter calibration mode
While in Measurement mode press the key. The display prompts you to enter the calibration
security code. Press the or key to set the calibration security code, then press the [ENTER]
key to confirm the calibration security code.
Press the or key to select one of the following calibration sub functions: Oxygen,
Temperature, Edit or Verify.
A flahing H in the top left hand corner shows the ongoing calibration process with a Hold
condition active.
7.3 Oxygen calibration
7.3.1 1-point calibration
Enter Calibration mode as described in section 7.2.
A O
2
sensor calibration is always a one point calibration either an Air (Slope) or a Zero (Offset)
calibration.
A one point slope calibration is done in air and a one point offset calibration is done at 0 ppb
O
2
. Press the [ENTER] key after selecting Slope or Offset.
Enter the value for Point 1 including a decimal point. O
2
is the value being measured by the
transmitter and sensor in the units set by the user. Press [ENTER] when this value is stable to
perform the calibration.
After the calibration the slope calibration factor S and the offset calibration factor Z are displayed.
Select „Yes" to save the calibration values and the successful Calibration is confirmed on the
display.