METTLER TOLEDO PS Shipping Scale Technical/Operators Manual
4.
Connect the power to the scale.
This prompt only appears if the
calibration jumper is removed.
5. Enter the setup mode by pressing and holding the UNITS key for up to
eight seconds until the message Setup is displayed. Release the UNITS
key.
6. Press the UNITS key several times until the Cal prompt appears, then
press ZERO to display YES or NO. Select Yes to calibrate or select No to
abort calibration. If Yes, press UNITS to begin the calibration process.
7. Press UNITS to select the displayed approval option.
8. Press UNITS until the Alt prompt appears, then press ZERO to display the
desired primary and secondary units. Select from the following options: (Not
an option on PS6L)
kg/kg
lb/lb
lb/kg
kg/lb
Calibration units are selected according to the scale build. For example, if
a scale has a kilogram build, select kg as the primary unit. Alternately, if
the scale has a pound build, select lb as the primary unit.
9. Press UNITS to select the displayed units option.
10. Press UNITS until the Build prompt appears, then press ZERO to display
the desired capacity/increment option. Capacity/increment options include:
If a different capacit
y
/increment option
is selected, the Cal prompt
automatically appears.
PS60/PS60U
Display Prompt Build
60-01
150-02
60-02
150-05
60-ni
200-ni
70-ni
60 x 0.01 kg
150 x 0.02 lb
60 x 0.02 kg
150 x 0.05 lb
60 x 0.01 / 0.02 kg Multi-Interval
200 x 0.05 / 0.1 lb Multi-Interval
70 x 0.02 / 0.05 kg Multi-Interval
PS6L
Note: The PS6L has only one Capacity/Increment build: 0-10 lb x 0.1 oz, 10-
70 x 0.2 oz, 70-150 lb x 0.5 oz.
PS30/PS3L
(08/04) 2-10