Chapter 4: Set Up Parameters and Calibration
4-1 (08/09)
TECHNICAL MANUAL
4 Setup Parameters and Calibration
This chapter discusses basic information and specific instructions for configuring each
program block of the PS scale, setting the operating parameters, and calibrating the PS
scale.
The scale’s interface program is shown in the quick overview map on the next page.
There are 17 functions or programming steps available, with two levels of access:
o Normal set up parameters that can be accessed by any operator
o Full access to all set up parameters with the removal of the metrology lock (IE:
calibration jumper or switch W1).
The functionality behind the Metrology lock is shown in the blue blocks on the program
overview map with a Lock Symbol .
For a quick, one-selection SET UP process without calibration, the PS scale menu has a
specific default setting for each of the major Shipping Carriers (FedEx, UPS, DHL NA,
Purolator, etc). Go directly to the END function (Step 17) and select the default for the
carrier and the scale will automatically input the other parameters. Please use one of the
default settings if you do not wish to customize the parameter settings and you are not
calibrating the scale with test weights.
Remove Cal Jumper
To access the full program, the metrology lock will need to be set to the off position by
removing the Calibration jumper W1. To remove W1:
Disconnect power to the scale, remove the platter, and remove the PCB cover:
o PS60, PS30, PS3L, and PS6L – Cover is located behind the display and under
the spider that holds the platter. Use 9/32 socket or adjustable wrench to remove
bolt.
o PS15 – Turn scale over and remove bolt on plastic cover. Use 9/32 socket or
adjustable wrench to remove bolt.
o PS90 – PCB is located under stainless steel connector cover behind the display.
Remove two screws on left and
right side of cover.
Inside the cover, there is a W1 (blue
plastic) jumper indicated by the red
dashed circle in photo. Remove jumper
off of both prongs and replace jumper
on only one prong so it doesn’t get lost.
Replace cover loosely and place platter
back on the scale. Connect the power
source to the scale.
Legal-for-Trade Application
Note: Accessing the calibration
switch will require breaking the
Weights and Measure Seal if
already applied. Breaking the
Weights and Measure seal will
require re-certification of the scale