4.3.5 TB3 – Analog pH / ORP sensors
pH / ORP sensors use 52 300 1XX series VP cables, or 10 001 XX02 series AS9 cables
(ORP only).
Pin no. Sensor wire color Function
1 Coax inner / transparent Glass
2 not used
3* Coax shield / red Reference
4* green / yellow, blue Solution GND / Shield
5 not used
6 white RTD ret / GND
7 RTD sense
8 green RTD
9 +5 V
grey (no connection)
Take care that AS9 cable and AK9 cable have the same configuration. So, if you want AS9 cable
with InPro 2000 and AK9 cable with InPro 3030 connect to TB3, do it as DPAS sensor.
Pin no. 1: Sensing (electrode).
Pin no. 3: Reference (Install jumper 3 to 4).
Terminal 4 and 6 are internally connected, either terminal can be used to connect a wire.
h
NOTE: * Install Jumper 3 to 4 when used without Solution Ground.
4.3.6 TB3 – Analog ISFET sensors
ISFET sensors use 52 300 40X series VP cables
Pin no. Sensor wire color Function
1 Coax inner / pink FET
2 not used
3 yellow Reference
4 green / yellow GND / Shield
5 not used
6 white RTD ret / GND
7 not used
8 grey RTD
9 brown +5 V
h
NOTE: Jumper 3 to 4 has to be installed
Terminal 4 and 6 are internally connected, either terminal can be used to connect a wire.