7.10 Edit sensor calibration constants
(only for analog sensor)
Enter Calibration mode as described in section 7.1 “Enter Calibration Mode” and select Edit, Edit
pH, Edit ISFET, or Edit mV.
All calibration constants for the selected sensor channel are displayed. Primary measurement
constants (p) are displayed on Line 3. Secondary measurement (temperature) constants (s)
for the sensor are displayed on Line 4.
The calibration constants can be changed in this menu.
Select Yes to save the new calibration values and the successful calibration is confirmed on
the display.
h
NOTE: Each time a new analog conductivity sensor is connected to the M400 Type 1, 2, 3
transmitter, it is necessary to enter the unique calibration data (cell constant and offset) located
on the sensor label.
7.11 Sensor verication
Enter Calibration mode as described in section 7.1. “Enter Calibration Mode” and select Verify.
The measured signal of the primary and the secondary measurement in electrical units
are shown. The meter calibration factors are used when calculating these values.
Press [ENTER] to exit from this display.