METTLER TOLEDO Weigh Module Systems Handbook
(12/99)4-10
NIST Handbook 44 legal-for-trade tolerances. The service/storage range is the
temperature range in which the load cell will operate without physical damage.
METTLER TOLEDO Load Cells
Compensated Range -10°C to +40°C (+14°F to +104°F)
Service/Storage Range -50°C to +85°C (-58°F to +185°F)
Table 4-3: Load Cell Temperature Specifications
In applications with high temperatures inside the tank, you can reduce thermal
conduction by placing insulation between the tank and the weigh modules. Use
insulating material with a compressive strength above 15,000 psi and thermal
conductivity ratings below 2.0 BTU-in/ft
2
/hr. The material must be able to withstand the
exposure temperature for prolonged periods without breaking down or deforming. Two
recommended FDA-approved materials are listed below:
Acetron® GP Acetal (Acetron is a registered trademark of DSM)
Continuous Service Temperature: 180°F
Heat Deflection Temperature at 264 psi: 220°F
Thermal Conductivity: 1.6 BTU inches/hour/foot
2
/°F
Coefficient of Thermal Linear Expansion: 5.4 × 10
-5
Compressive Strength: 15,000 psi
Ultem 1000 Polyetherimide (PEI)
Continuous Service Temperature: 340°F
Heat Deflection Temperature at 264 psi: 392°F
Thermal Conductivity: 0.9 BTU inches/hour/foot
2
/°F
Coefficient of Thermal Linear Expansion: 3.1 × 10
-5
Compressive Strength: 22,000 psi
Moisture and Corrosion
Moisture or corrosive material on a weigh module can affect the life of the load cells.
Debris, such as leaves and dirt, accumulated in and around weigh modules can also
cause problems. There are several steps you can take to minimize the potential for
moisture and corrosion problems:
Provide adequate drainage away from the weigh modules.
Keep weigh modules clear of snow that will melt and introduce moisture into the
system.
Do not use tanks with flat tops that catch and retain water, snow, leaves, or other
debris that will add uncompensated weight to the system.
Hose down the tanks regularly to clean accumulated debris.
Keep cables clean and in good condition. Broken cables or worn cable sheathing
can allow water to enter and cause corrosion.