Mettler Toledo M400/2(X)H Spezifikationen Seite 37

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 111
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 36
Transmitter M400/2(X)H, M400G/2XH 37
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
Enter the value for Point 1 including a decimal point and units. The value in the second text line
is the value being measured by the transmitter and sensor in the units selected by the user.
Press [ENTER] when this value is stable to perform the calibration.
After the calibration the cell multiplier or slope calibration factor ”M” i.e. cell constant and the
Adder or offset calibration factor ”A” are displayed.
In case of a successful calibration, the calibration values are stored in the cal history* and taken
over (Adjust), stored in the cal history* and not taken over (Calibrate) or discarded (Abort).
* only available with ISM sensor. The values will be stored in the sensor.
If ”Adjust” or ”Calibrate” are chosen, the message ”Calibration successful” is displayed. In any
case you will get the message ”Re-install sensor” and ”Press ENTER” on the display. After press-
ing ”ENTER” the M400 returns to the measuring mode.
7.2.2 Two-point sensor calibration
(four electrode sensors only)
(Display reflects typical Conductivity sensor calibration)
Enter Conductivity Sensor Calibration mode as described in section 7.1 “Enter Calibration Mode”
and choose one of the compensation modes (see section 7.2 “Conductivity calibration for two
or four electrode sensors”).
Select 2 point calibration and press [ENTER].
Place the electrode into the first reference solution.
CAUTION: Rinse sensors with a high-purity water solution between calibration points to prevent
contamination of the reference solutions.
Enter the value for Point 1 including a decimal point and units. The value in the second text line
is the value being measured by the transmitter and sensor in the units selected by the user.
Press [ENTER] when this value is stable and place the electrode into the second reference
solution.
Enter the value for Point 2 including a decimal point and units. The value in the second text line
is the value being measured by the transmitter and sensor in the units selected by the user.
Press [ENTER] when this value is stable to perform the calibration.
After the calibration of the cell multiplier or slope calibration factor ”M” i.e. cell constant and the
adder or offset calibration factor ”A” are displayed.
In case of a successful calibration, the calibration values are stored in the cal history* and taken
over (Adjust), stored in the cal history* and not taken over (Calibrate) or discarded (Abort).
* only available with ISM sensor. The values will be stored in the sensor.
If ”Adjust” or ”Calibrate” are chosen, the message ”Calibration successful” is displayed. In any
case you will get the message ”Re-install sensor” and ”Press ENTER” on the display. After press-
ing ”ENTER” the M400 returns to the measuring mode.
Seitenansicht 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 110 111

Kommentare zu diesen Handbüchern

Keine Kommentare