8.2.3.2 Concentration table
If during the channel setup (see chapter 8.2.1 “Channel setup”) the parameter conductivity has
been chosen or an four-electrode conductivity sensor based on ISM technology is connected to
the transmitter, a concentration table can be defined.
To specify customers-specific solutions, up to 9 concentration values can be edited in a matrix
together with up to 9 temperatures. To do so the desired values are edited under the concentra-
tion table menu. Furthermore the conductivity values for the according temperature and concen-
tration values are edited.
For doing the settings the menu “Concentration Table”, that will be displayed, has to be chosen.
(see chapter 8.2.3 “Parameter related settings”).
Define the desired unit.
Press [ENTER]
h
NOTE: Refer to section 8.2.1 “Channel Setup” to choose the unit used in the display.
Enter the amount of desired temperature points (Temp Point) and Concentration Points.
Press [ENTER]
Enter the values for the different concentrations (ConcentrationX).
Press [ENTER]
Enter the value of the 1st temperature (Temp1) and the value for the conductivity which belongs
to the first concentration at this temperature.
Press [ENTER]
Enter the value for the conductivity which belongs to the second concentration at the first temper-
ature and press [ENTER] etc..
After entering all conductivity values, that belong to the different concentrations at the first tem-
perature point, enter in the same way the value of the 2nd temperature point (Temp2) and the
conductivity value which belongs at the second temperature to the first concentration. Press
[ENTER] and go on in the same way for the next concentration points as described for the first
temperature point.
Enter in this way the values at every temperature point. After entering the last value, pressing
[ENTER] again will bring up the Save Changes dialog. Selecting No will discard the entered val-
ues and return to the measurement display screen, selecting Yes will save changes made.
h
NOTE: The values for the temperature have to increase from Temp1 to Temp2 to Temp3 etc.. The
values for the concentration have to increase from Concentration1 to Concentration2 to
Concentration3 etc..