9.2.1 Changing passwords
See Section 9.3 on how to enter the passwords menu. Select Change Administrator or Change
Operator and set the new password.
Press the [ENTER] key and confirm the new password. Press [ENTER] again to bring up the
Save Changed dialog.
9.2.2 Conguring menu access for operator
See 9.3 on how to enter the passwords Menu. Select Configure Operator to configure the access
list for the operator. It is possible to assign / deny rights to the following menus: Cal Key, Quick
Setup, Configuration, System, PID Setup and Service.
Choose either Yes or No to give / deny access to the above menus and press [ENTER] to
advance to the next items. Pressing the [ENTER] key after configuring all menus will bring up
the Save Changes dialog. Selecting No will discard the entered values, selecting Yes will make
the entered values the current ones.
9.3 Set / Clear lockout
(PATH: Menu / System / Set / Clear Lockout)
This menu enables / disables the lockout functionality of the transmitter. The user will be asked
for a password before being allowed into any menus if the lockout functionality is enabled.
The lockout-menu is protected: Enter the administrator or operator password and select YES to
enable or NO to disable the lockout functionality. Pressing the [ENTER] key after the selection
will bring up the Save Changes dialog. Selecting No will discard the entered value, selecting Yes
will make the entered value the current one.
9.4 Reset
(PATH: Menu / System / Reset)
This menu allows access to the following options:
Reset System, Reset Meter Cal, Reset Analog Cal.