Mettler Toledo M400/2(X)H Spezifikationen Seite 43

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 111
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 42
Transmitter M400/2(X)H, M400G/2XH 43
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
The display shows the buffer the transmitter has recognized (Point 1) and the measured value.
As soon as the stabilisation criteria have been fulfilled stabilisation criteria have been fulfilled,
the display changes and prompts you to place the electrode in the second buffer.
Place the electrode in the second buffer solution and press the [ENTER] key to go on with the
calibration.
The display shows the second buffer the transmitter has recognized (Point 2) and the measured
value.
As soon as the stabilisation criteria have been fulfilled the display changes to show the slope
calibration factor S and the offset calibration factor Z.
In case of a successful calibration, the calibration values are stored in the cal history* and taken
over (Adjust), stored in the cal history* and not taken over (Calibrate) or discarded (Abort).
* only available with ISM sensor. The values will be stored in the sensor.
If “Adjust” or “Calibrate” are chosen, the message “Calibration successful” is displayed. In any
case you will get the message “Re-install sensor” and “Press ENTER” on the display. After press-
ing “ENTER” the M400 returns to the measuring mode.
7.4.2.2 Manual Mode
Place the electrode in the first buffer solution. The display shows the buffer the transmitter has
recognized (Point 1) and the measured value. Press [ENTER] to proceed.
Place the transmitter in the second buffer solution. The display shows the buffer the transmitter
has recognized (Point 2) and the measured value. Press [ENTER] to proceed.
The display shows the slope calibration factor S and the offset calibration factor Z.
In case of a successful calibration, the calibration values are stored in the cal history* and taken
over (Adjust), stored in the cal history* and not taken over (Calibrate) or discarded (Abort).
If “Adjust” or “Calibrate” are chosen, the message “Calibration successful” is displayed. In any
case you will get the message “Re-install sensor” and “Press ENTER” on the display. After press-
ing “ENTER” the M400 returns to the measuring mode.
Seitenansicht 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 110 111

Kommentare zu diesen Handbüchern

Keine Kommentare